Ionization Dose to Water: A More Accurate Measure for Radiation Therapy Doses


Основні поняття
This paper proposes a new dosimetry concept, "ionization dose to water," to improve the accuracy and relevance of radiation therapy dose measurement, particularly for proton and ion beams, by focusing on ionization rather than heat as the primary measurement.
Анотація
  • Bibliographic Information: Kanematsu, N. (2024). Conceptualization of ionization dose to water to quantify radiation therapy doses. arXiv preprint arXiv:2410.21948v1.
  • Research Objective: This study aims to conceptualize and formulate a new dosimetry method based on ionization, termed "ionization dose to water," to improve the accuracy of radiation therapy dose measurement, especially for proton and ion beams.
  • Methodology: The study leverages the existing International Code of Practice (ICP) for ionization-chamber dosimetry and proposes a dosimetric procedure that eliminates the need for W-value correction for all beams and stopping-power-ratio correction for proton and ion beams. The authors design water-equivalent gas (WEG) mixtures for use in ionization chambers to minimize uncertainties. Reference dosimetry simulations are performed for various beam types to demonstrate the feasibility of the proposed method.
  • Key Findings: The proposed ionization dose measurement is largely equivalent to absorbed dose for photon and electron beams. For proton and ion beams, using WEG mixtures significantly reduces dosimetric uncertainties to 0.7% and 1.0%, respectively, compared to conventional methods. The study also identifies potential challenges with different WEG mixtures, such as leakage with helium and flammability with methane and ethane.
  • Main Conclusions: The ionization dose to water, with its minimal reliance on beam quality corrections, offers a more accurate and relevant measure for radiation therapy doses, particularly for proton and ion beams. This concept has the potential to improve dosimetry accuracy in non-reference conditions and for complex clinical beams.
  • Significance: This research introduces a novel dosimetry concept with the potential to significantly improve the accuracy and consistency of radiation therapy dose delivery, ultimately leading to better treatment outcomes.
  • Limitations and Future Research: Further research is needed to determine the optimal WEG mixture and chamber design for practical implementation. Additionally, detailed Monte Carlo simulations are necessary to establish comprehensive chamber perturbation factors for the ionization dose.
edit_icon

Налаштувати зведення

edit_icon

Переписати за допомогою ШІ

edit_icon

Згенерувати цитати

translate_icon

Перекласти джерело

visual_icon

Згенерувати інтелект-карту

visit_icon

Перейти до джерела

Статистика
The relative uncertainty in dose measurement using conventional methods is excessively large for proton (1.4%) and ion beams (2.4%) compared to photon (0.62%) and electron beams (0.68%). The largest contribution to uncertainty for proton and ion beams comes from the WairQ value (mean energy expended in air per ion pair formed). The proposed ionization dose measurement method, using water-equivalent gas, reduces dosimetric uncertainties to 0.7% and 1.0% for proton and ion beams, respectively.
Цитати
"Absorbed dose essentially measures heat per mass by temperature rise in reference calorimetry. However, since it is ionization that induces chemical reactions for cell damage in radiation therapy […], it is more direct and natural to measure dose by ionization rather than by heat […]." "For dosimetry in water as reference medium for its abundance in biological tissues, it is also more direct and natural to use water or water-equivalent material for the dosimeter […]." "The ionization dose with reduced beam quality corrections provides dose representations of improved accuracy and relevance to radiation therapy, particularly for proton and ion beams."

Ключові висновки, отримані з

by Nobuyuki Kan... о arxiv.org 10-30-2024

https://arxiv.org/pdf/2410.21948.pdf
Conceptualization of ionization dose to water to quantify radiation therapy doses

Глибші Запити

How might the adoption of ionization dose impact the calibration and standardization of radiation therapy equipment and procedures?

The adoption of ionization dose (DI) could bring about significant changes in the calibration and standardization of radiation therapy equipment and procedures, particularly for proton and ion beam therapy. Here's how: Shift from Absorbed Dose to Ionization Dose: The current paradigm relies on absorbed dose to water (D) as the primary quantity. A shift to DI would necessitate a reevaluation of existing calibration protocols and the development of new standards specifically for DI. Calibration Procedures: Currently, ionization chambers are calibrated against a reference ⁶⁰Co γ-ray beam to determine the dosimeter calibration coefficient (NDQ0). With DI, this calibration process would need to account for the mean ionization energy fraction (ιQ0) of the chamber gas. This might involve establishing reference values for ιQ0 for different gases and beam qualities. Water-Equivalent Gas Chambers: The development and widespread adoption of water-equivalent gas (WEG) ionization chambers would be crucial for the practical implementation of DI-based dosimetry, especially for proton and ion beams. This would require standardized procedures for WEG chamber manufacturing, gas mixing, and quality assurance. Beam Quality Correction Factors: The paper highlights the challenges associated with the beam quality correction factor (kQ/Q0) for proton and ion beams. With DI, the dependence on WairQ/Q0 is eliminated, potentially simplifying the correction factors. However, accurate determination of the remaining perturbation factors (pchQ/Q0) for WEG chambers would be essential. Standardization Bodies: International organizations like the IAEA and ICRU would play a key role in defining new standards, protocols, and recommendations for DI-based dosimetry. This would involve collaboration among physicists, dosimetrists, and manufacturers. Overall, the transition to ionization dose would require a concerted effort from the radiation therapy community to establish new calibration and standardization procedures. However, the potential benefits in terms of accuracy, relevance to biological effects, and simplification of beam quality corrections make it a promising avenue for future research and development.

Could the focus on ionization over heat potentially lead to an underestimation of dose in situations where other energy deposition mechanisms are significant?

Yes, focusing solely on ionization dose (DI) could potentially lead to an underestimation of the biologically effective dose in situations where other energy deposition mechanisms, besides ionization, play a significant role. Non-Ionizing Energy Deposition: Absorbed dose (D) accounts for all energy deposited in a medium, including contributions from ionization, excitation, and other processes. DI, by definition, only considers the energy expended specifically on ionization. While ionization is the dominant mechanism for radiation-induced biological damage, other processes can still contribute. Low-Energy Photons and Neutrons: In cases involving low-energy photons or neutrons, a significant portion of the energy deposition might occur through mechanisms like Compton scattering or elastic collisions, which may not directly lead to ionization in the dosimeter. This could result in an underestimation of the total energy deposited and potentially the biological effect. High-LET Radiation: For high-linear energy transfer (LET) radiation like heavy ions, the density of ionization events along the particle track is much higher. While DI would capture the ionization component accurately, it might not fully reflect the increased biological effectiveness associated with the dense ionization pattern. Biological Effectiveness: It's crucial to remember that DI, while more directly related to ionization, is not a direct measure of biological effect. The biological response to radiation depends on various factors, including radiation type, dose rate, and tissue sensitivity. Therefore, while DI offers advantages in terms of accuracy and relevance to ionization events, it's essential to consider its limitations. In situations where other energy deposition mechanisms are significant, relying solely on DI could lead to an underestimation of the biologically effective dose. A comprehensive approach considering both ionization and other energy deposition processes, along with biological effectiveness factors, would be necessary for accurate dose assessment and treatment planning.

If we can measure and quantify the biological effects of radiation more directly, how might this change our approach to radiation therapy planning and delivery?

The ability to directly measure and quantify the biological effects of radiation would revolutionize radiation therapy planning and delivery, leading to more personalized and effective treatments with potentially fewer side effects. Here's how: Personalized Treatment Planning: Currently, treatment plans are based on dose distributions calculated to achieve tumor control while sparing healthy tissues. Direct measurement of biological effects would enable the creation of "biological dose" maps, allowing for optimization of the treatment plan based on the predicted individual patient response. Normal Tissue Sparing: By precisely quantifying the biological impact on healthy tissues, we could tailor treatment plans to minimize long-term side effects. This is particularly crucial for organs at risk (OARs) with varying radiosensitivities. Real-Time Dose Painting: Imagine a scenario where we can monitor the biological response of tumors and healthy tissues during treatment delivery. This real-time feedback would allow for dose adjustments "on the fly," ensuring optimal tumor control while minimizing damage to surrounding tissues. Adaptive Radiation Therapy: Biological measurements could be integrated into adaptive radiation therapy (ART) workflows. By assessing the biological response at different treatment fractions, we could adapt the plan to account for tumor shrinkage, changes in patient anatomy, or variations in radiosensitivity. New Treatment Modalities: A deeper understanding of biological effects might pave the way for the development of novel treatment modalities. For instance, we could explore targeted therapies that enhance the radiosensitivity of tumor cells while protecting healthy tissues. However, several challenges need to be addressed: Measurement Techniques: Developing accurate, reliable, and real-time methods for measuring biological effects in vivo remains a significant hurdle. Biological Variability: Biological responses to radiation can vary greatly between individuals due to genetic factors, lifestyle, and other medical conditions. Accounting for this variability in treatment planning would be crucial. Ethical Considerations: The use of new biological markers and personalized treatments would raise ethical considerations regarding patient privacy, informed consent, and access to these advanced technologies. Despite these challenges, the ability to directly measure and quantify biological effects holds immense promise for the future of radiation therapy. It has the potential to transform cancer treatment, making it more precise, personalized, and effective.