Accuracy and Repeatability Assessment of a Parallel Robot for Personalized Minimally Invasive Surgical Applications


Core Concepts
The paper presents the experimental assessment of the accuracy and repeatability of a parallel robot (PARA-SILSROB) developed for personalized minimally invasive surgical applications.
Abstract

The paper discusses the development and experimental evaluation of the PARA-SILSROB parallel robot, which was designed to address the limitations of current minimally invasive surgical techniques. The robot is intended to control laparoscopic cameras and active surgical instruments during single-incision laparoscopic surgery (SILS).

The authors conducted experiments to assess the accuracy and repeatability of the PARA-SILSROB robot using two methods:

  1. Optical motion tracking: The accuracy was evaluated by comparing the trajectory of the robot's end-effector, measured using the OptiTrack system, with the trajectory generated using the robot's kinematic model. The root-mean-square error (RMSE) between the measured and modeled trajectories was found to be 0.3 mm.

  2. Coordinate measuring: The repeatability was determined by repeatedly moving the robot's mobile platform to a predefined point and measuring its position using a Stinger II measuring arm. The repeatability standard deviation for the entire trajectory was 0.18 mm.

The results demonstrate that the PARA-SILSROB robot meets the accuracy and repeatability requirements for surgical applications, which are typically in the range of 1-2 mm. The authors note that the accuracy and repeatability of the robot are influenced by factors such as the limitations of the optical tracking system and the elasticity of the tissue during actual surgical procedures.

The paper concludes by highlighting the promising results of the first iteration of the PARA-SILSROB robot and plans for future development, including extended studies on different trajectories and experimental tests on human phantoms in relevant medical conditions.

edit_icon

Customize Summary

edit_icon

Rewrite with AI

edit_icon

Generate Citations

translate_icon

Translate Source

visual_icon

Generate MindMap

visit_icon

Visit Source

Stats
The accuracy of the PARA-SILSROB robot was determined using the root-mean-square error (RMSE) between the measured trajectory from the OptiTrack system and the trajectory generated using the robot's kinematic model. The RMSE values were 0.2866 mm for the X-axis, 0.4052 mm for the Y-axis, and 0.1161 mm for the Z-axis. The repeatability of the PARA-SILSROB robot was determined using the Stinger II measuring arm. The repeatability standard deviation for the entire trajectory was 0.18 mm.
Quotes
"The accuracy of the system for this specific trajectory was determined using the RMSE between the RMSE of OptiTrack and the RMSE of values extracted from the experimental model during the experimental run and the result is 0.3 mm, this accuracy is influenced by the OptiTrack system limitation and the light sources, according to the technical requirement of this system the accuracy is 0.1 mm in ideal conditions." "The repeatability for point 1 was 0.37 mm, for point 2 was 0.25mm, for point 3 was 0.31mm and for point 4 was 0.23mm. The repeatability standard deviation for the entire robotic system on the trajectory defined above is 0.18."

Deeper Inquiries

How can the accuracy and repeatability of the PARA-SILSROB robot be further improved, considering the limitations of the optical tracking system and the challenges posed by the elasticity of tissue during actual surgical procedures?

To enhance the accuracy and repeatability of the PARA-SILSROB robot, several strategies can be implemented. Firstly, improving the calibration process of the optical tracking system can help reduce errors in tracking the robot's movements. This can involve refining the placement of markers, optimizing the camera positions, and minimizing external factors that may interfere with the tracking system's performance. Additionally, incorporating redundant sensing mechanisms, such as integrating force sensors or tactile feedback systems, can provide supplementary data to validate the robot's position and ensure precise control. By combining multiple sensing modalities, the system can compensate for inaccuracies in any single measurement method, thereby enhancing overall accuracy. Moreover, developing advanced algorithms for real-time error correction and compensation can mitigate the impact of tissue elasticity during surgical procedures. These algorithms can account for dynamic changes in tissue properties and adjust the robot's movements accordingly to maintain accuracy and repeatability. Furthermore, exploring novel materials and designs for the robot's end-effectors that are more compatible with tissue properties can help minimize errors caused by tissue deformation. By optimizing the physical characteristics of the robot's tools, such as stiffness and compliance, the system can better adapt to the challenges posed by tissue elasticity.

What additional factors, beyond accuracy and repeatability, should be considered when evaluating the performance and suitability of the PARA-SILSROB robot for personalized minimally invasive surgical applications?

In addition to accuracy and repeatability, several other factors are crucial when assessing the performance and suitability of the PARA-SILSROB robot for personalized minimally invasive surgical applications. These factors include: Safety: Ensuring the robot's design and control mechanisms prioritize patient safety and minimize the risk of intraoperative complications. Speed and Efficiency: Evaluating the robot's speed in executing surgical tasks and its overall efficiency in reducing procedure duration while maintaining precision. Ease of Use: Assessing the user interface, ergonomics, and intuitiveness of the control system to facilitate seamless operation by surgeons. Scalability: Considering the robot's adaptability to different surgical procedures and its potential for customization based on individual patient needs. Cost-effectiveness: Analyzing the overall cost of implementing the robot in surgical settings, including initial investment, maintenance, and training expenses. Integration with Existing Systems: Ensuring compatibility with existing surgical equipment and infrastructure to facilitate seamless integration into clinical workflows. Clinical Outcomes: Monitoring and evaluating the impact of the robot on patient outcomes, such as postoperative recovery time, complication rates, and overall surgical success.

How can the insights gained from the development and evaluation of the PARA-SILSROB robot be applied to the design and optimization of other parallel robotic systems for medical applications?

The insights obtained from the development and evaluation of the PARA-SILSROB robot can be instrumental in enhancing the design and optimization of other parallel robotic systems for medical applications. Some key applications include: Kinematic Modeling: Utilizing the kinematic models and trajectory planning algorithms developed for the PARA-SILSROB to optimize the motion control and path planning of other parallel robotic systems in medical settings. Sensing and Tracking: Implementing the lessons learned from the accuracy and repeatability assessments to enhance the sensing and tracking capabilities of other robotic platforms, improving overall performance and reliability. Human-Robot Interaction: Incorporating the master-slave control concept and ergonomic considerations from the PARA-SILSROB into the design of other robotic systems to enhance human-robot interaction and user experience. Safety Features: Integrating safety mechanisms and error detection algorithms derived from the PARA-SILSROB evaluation to ensure the safe operation of parallel robotic systems in medical environments. Customization and Adaptability: Applying the principles of customization and adaptability observed in the PARA-SILSROB to tailor other robotic systems for specific surgical procedures and patient requirements, enhancing versatility and applicability. By leveraging the knowledge and experiences gained from the PARA-SILSROB project, designers and researchers can advance the development of parallel robotic systems for a wide range of medical applications, ultimately improving patient care and surgical outcomes.
0
star