Bibliographic Information: Sharafi, A., Mickevicius, N.J., Baboli, M., Nencka, A.S., & Koch, K.M. (2024). Variable Resolution Sampling and Deep Learning Image Recovery for Accelerated Multi-Spectral MRI Near Metal Implants. arXiv preprint arXiv:2410.23329v1.
Research Objective: To develop and validate a variable resolution sampling and deep learning reconstruction framework for multi-spectral MRI that reduces scan times without compromising image quality, specifically for imaging near metal implants.
Methodology: This retrospective study utilized proton density-weighted 1.5T MSI knee and hip data from patients with metal implants. A novel spectral undersampling scheme, acquiring full k-space data for half the spectral bins and only auto-calibration signal (ACS) data for the remaining bins, was employed to achieve a 40% reduction in scan time. A multi-channel 2D U-Net model was trained to reconstruct full-resolution images from the undersampled data. Image quality was evaluated using SSIM, PSNR, and RESI metrics, comparing the deep learning reconstructions to conventionally reconstructed images and fully sampled references.
Key Findings: Deep learning reconstructions of the undersampled data showed significantly higher SSIM and PSNR values compared to conventional reconstructions, indicating improved image quality and reduced reconstruction error. The edge sharpness of deep learning reconstructed images was comparable to that of fully sampled references, demonstrating the method's ability to preserve anatomical details.
Main Conclusions: The proposed variable resolution sampling and deep learning reconstruction approach can potentially enhance MRI examinations near metal implants by reducing scan times without sacrificing image quality. This could lead to faster exams, higher resolution imaging, and improved diagnostic capabilities.
Significance: This research offers a promising solution to the challenge of long scan times associated with multi-spectral MRI, particularly beneficial for imaging near metal implants where artifacts are prevalent. The use of deep learning allows for efficient and accurate image recovery from undersampled data, potentially improving the clinical workflow and expanding the applications of MSI.
Limitations and Future Research: The study was retrospective and focused on quantitative image quality metrics. Future research should include prospective studies with a comprehensive reader study to evaluate the diagnostic accuracy and clinical impact of this approach. Additionally, investigating the effects of different echo train view ordering in prospectively accelerated MSI and exploring alternative deep learning architectures could further optimize the method.